Technical Specifications: | ||
Parameters | WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PDW, PCT CBC+DIFF for Dog, Cat, Horse, Rat, Mouse, Rabbit, Cow and Monkey Lymph#, Mon#, Gran#, Lymph%, Mon%, Gran%, Eos% (Special for Dog and Cat) Histograms for WBC, RBC and PLT | |
Principles | Electrical impedance method for cell counting and cyanide free method for Hemoglobin | |
Carryover | WBC, RBC, HGB≤0.5%, PLT≤1.0% | |
Sample Volume | Prediluted: 20μL Whole Blood: 13μL | |
Throughput | 25 samples per hour | |
Data storage | 10,000 results with histograms | |
Display | Color LCD display resolution: 800x600 | |
Input/Output | Rs232x2, 1 parallel printer (optional), 1 barcode scanner (optional), 1 keyboard interface | |
Printout | Thermal recorder, 50mm width paper, multiple printout formats, external printer (optional) | |
Operating Environment | Temperature: 15℃~30℃ Humidity: 30%~85% | |
Power Requirement | AC 100-240V, 50/60Hz | |
Dimensions | 322mm (W)x437mm(H)x386mm(D) | |
Weight | Less than 23kg |
Performance | ||
Parameter | Linearity Range | Precision (CV%) |
WBC (109 /L) | 0.4~99.9 | 3.0 (7.5~15.0) |
RBC (1012 /L) | 0.20~16.99 | 2.0 (3.50~6.00) |
HGB (g/L) | 3-250 | 2.0 (120~180) |
MCV (fL) | 2.0 (40~110.0) | |
PLT (109 /L) | 10-2999 | 6.0 (200~500) |
Contact Person : | Ginny |
---|